Module 3 - Introduction to Circuit Protection, Control, and Measurement
Navy Electricity and Electronics Training Series (NEETS)
Chapter 1:  Pages 1-31 through 1-40

Module 3 −− Introduction to Circuit Protection, Control, and Measurement

Pages i, 1−1, 1−11, 1−21, 1−31, 1−41, 1−51, 1−61, 1−71, 2−1, 2−11, 1−21, 2−31, 2−41, 3−1, 3−11, 3−21, 3−31, AI−1, AII−1, AIII−1, IV−1, Index

 

 

Reading a voltmeter at various ranges - RF Cafe

Figure 1-30. - Reading a voltmeter at various ranges.

In figure 1-30(A) the meter is in the 1000-volt range. The pointer is barely above the 0 position. It is not possible to accurately read this voltage. In figure 1-30(B) the meter is switched to the 250 volt range. From the pointer position it is possible to approximate the voltage as 20 volts. Since this is well below the next range, the meter is switched, as in figure 1-30(C). With the meter in the 50-volt range, it is possible to read the voltage as 22 volts. Since this is more than the next range of the meter (10 volts), the meter would not be switched to the next (lower) scale.

Q34.    How is it possible to use a current sensitive meter movement to measure voltage?

Q35.    What is voltmeter sensitivity?

Q36.    What method is used to allow a voltmeter to have several ranges?

Q37.    Why should you always use the highest range when connecting a voltmeter to a circuit?

ELECTROSTATIC METER MOVEMENT

The final meter movement covered in this chapter is the ELECTROSTATIC METER MOVEMENT. The other meter movements you have studied all react to current, the electrostatic meter movement reacts to voltage.

The mechanism is based on the repulsion of like charges on the plates of a capacitor. The electrostatic meter movement is actually a large variable capacitor in which one set of plates is allowed to

1-31

move. The movement of the plates is opposed by a spring attached to the plates. a pointer that indicates the value of the voltage is attached to these movable plates. As the voltage increases, the plates develop more torque. To develop sufficient torque, the plates must be large and closely spaced. a very high voltage is necessary to provide movement, therefore, electrostatic voltmeters are used only for High Voltage measurement.

VOLTMETER Safety PRECautionS

Just as with ammeters, voltmeters require safety precautions to prevent injury to personnel and damage to the voltmeter or equipment. The following is a list of the MINIMUM safety precautions for using a voltmeter.

·   Always connect voltmeters in parallel.

·   Always start with the highest range of a voltmeter.

·   Deenergize and discharge the circuit completely before connecting or disconnecting the voltmeter.

·   In dc voltmeters, observe the proper circuit polarity to prevent damage to the meter.

·   Never use a dc voltmeter to measure ac voltage.

·   Observe the general safety precautions of electrical and electronic devices.

Q38.    What type of meter movement reacts to voltage rather than current?

Q39.    What is the only use for the voltage sensitive meter movement?

Q40.    List the six safety precautions for the use of voltmeters.

OHMMETERS

The two instruments most commonly used to check the continuity (a complete circuit), or to measure the resistance of a circuit or circuit element, are the OHMMETER and the MEGGER (megohm meter). The ohmmeter is widely used to measure resistance and check the continuity of electrical circuits and devices. Its range usually extends to only a few megohms. The megger is widely used for measuring insulation resistance, such as between a wire and the outer surface of the insulation, and insulation resistance of cables and insulators. The range of a megger may extend to more than 1,000 megohms.

The ohmmeter consists of a dc ammeter, with a few added features. The added features are:

1.    a dc source of potential (usually a 3-volt battery)

2.    One or more resistors (one of which is variable)

3.    a simple ohmmeter circuit is shown in figure 1-31.

The ohmmeter's pointer deflection is controlled by the amount of battery current passing through the moving coil. Before measuring the resistance of an unknown resistor or electrical circuit, the test leads of the ohmmeter are first shorted together, as shown in figure 1-31. With the leads shorted, the meter is calibrated for proper operation on the selected range. While the leads are shorted, meter current is maximum and the pointer deflects a maximum amount, somewhere near the zero position on the ohms

1-32

scale. Because of this current through the meter with the leads shorted, it is necessary to remove the test leads when you are finished using the ohmmeter. If the leads were left connected, they could come in contact with each other and discharge the ohmmeter battery. When the variable resistor (rheostat) is adjusted properly, with the leads shorted, the pointer of the meter will come to rest exactly on the zero position. This indicates ZERO Resistance between the test leads, which, in fact, are shorted together. The zero reading of a series-type ohmmeter is on the right-hand side of the scale, where as the

zero reading for an ammeter or a voltmeter is generally to the left-hand side of the scale. (There is another type of ohmmeter which is discussed a little later on in this chapter.) When the test leads of an ohmmeter are separated, the pointer of the meter will return to the left side of the scale. The interruption of current and the spring tension act on the movable coil assembly, moving the pointer to the left side (∞) of the scale.

A simple ohmmeter circuit - RF Cafe

Figure 1-31. - a simple ohmmeter circuit.

USING The OHMMETER

After the ohmmeter is adjusted for zero reading, it is ready to be connected in a circuit to measure resistance. a typical circuit and ohmmeter arrangement is shown in figure 1-32.

1-33

Measuring circuit resistance with an ohmmeter - RF Cafe

Figure 1-32. - Measuring circuit resistance with an ohmmeter.

The power switch of the circuit to be measured should always be in the ofF position. This prevents the source voltage of the circuit from being applied across the meter, which could cause damage to the meter movement.

The test leads of the ohmmeter are connected in series with the circuit to be measured (fig. 1-32). This causes the current produced by the 3-volt battery of the meter to flow through the circuit being tested. Assume that the meter test leads are connected at points a and b of figure 1-32. The amount of current that flows through the meter coil will depend on the total resistance of resistors R1  and R2, and the resistance of the meter. Since the meter has been preadjusted (zeroed), the amount of coil movement now depends solely on the resistance of R1and R2. The inclusion of R1  and R2  raises the total series resistance, decreasing the current, and thus decreasing the pointer deflection. The pointer will now come to rest at a scale figure indicating the combined resistance of R1  and R2. If R1  or R2, or both, were replaced with a resistor(s) having a larger value, the current flow in the moving coil of the meter would be decreased further. The deflection would also be further decreased, and the scale indication would read a still higher circuit resistance. Movement of the moving coil is proportional to the amount of current flow.

OHMMETER RANGES

The amount of circuit resistance to be measured may vary over a wide range. In some cases it may be only a few ohms, and in others it may be as great as 1,000,000 ohms (1 megohm). To enable the meter to indicate any value being measured, with the least error, scale multiplication features are used in most ohmmeters. For example, a typical meter will have four test lead jacks-Common, R x 1, R x 10, and R x 100. The jack marked Common is connected internally through the battery to one side of the moving coil of the ohmmeter. The jacks marked R x 1, R x 10, and R x 100 are connected to three different size resistors located within the ohmmeter. This is shown in figure 1-33.

1-34

An ohmmeter with multiplication jacks - RF Cafe

Figure 1-33. - An ohmmeter with multiplication jacks.

Some ohmmeters are equipped with a selector switch for selecting the multiplication scale desired, so only two test lead jacks are necessary. Other meters have a separate jack for each range, as shown in figure 1-33. The range to be used in measuring any particular unknown resistance (Rx  in figure 1-33) depends on the approximate value of the unknown resistance. For instance, assume the ohmmeter in figure 1-33 is calibrated in divisions from 0 to 1,000. If Rx  is greater than 1,000 ohms, and the R x 1 range is being used, the ohmmeter cannot measure it. This occurs because the combined series resistance of resistor R x 1 and Rx  is too great to allow sufficient battery current to flow to deflect the pointer away from infinity (∞). (Infinity is a quantity larger than the largest quantity you can measure.) The test lead would have to be plugged into the next range, R x 10. With this done, assume the pointer deflects to indicate 375 ohms. This would indicate that Rx  has 375 ohms x 10, or 3,750 ohms resistance. The change of range caused the deflection because resistor R x 10 has about 1/10 the resistance of resistor R x 1. Thus, selecting the smaller series resistance permitted a battery current of sufficient amount to cause a useful pointer deflection. If the R x 100 range were used to measure the same 3,750-ohm resistor, the pointer would deflect still further, to the 37.5-ohm position. This increased deflection would occur because resistor R x 100 has about 1/10 the resistance of resistor R x 10.

The foregoing circuit arrangement allows the same amount of current to flow through the meter's moving coil whether the meter measures 10,000 ohms on the R x 10 scale, or 100,000 ohms on the R x 100 scale.

It always takes the same amount of current to deflect the pointer to a certain position on the scale (midscale position for example), regardless of the multiplication factor being used. Since the multiplier resistors are of different values, it is necessary to ALWAYS "zero" adjust the meter for each multiplication fact or selected.

You should select the multiplication factor (range) that will result in the pointer coming to rest as near as possible to the midpoint of the scale. This enables you to read the resistance more accurately, because the scale readings are more easily interpreted at or near midpoint.

1-35

Q41.    What electrical quantity is measured by an ohmmeter?

Q42.    What other measurement can an ohmmeter make?

Q43.    How is a series-type ohmmeter connected to the circuit being measured?

Q44.    What is used to provide the ohmmeter with several ranges?

Q45.    What area of an ohmmeter scale should be used when measuring circuits?

Shunt OHMMETER

The ohmmeter described to this point is known as a series ohmmeter, because the resistance to be measured is in series with the internal resistors and the meter movement of the ohmmeter. Another type of ohmmeter is the Shunt OHMMETER. In the shunt ohmmeter, the resistance to be measured shunts (is in parallel with) the meter movement of the ohmmeter. The most obvious way to tell the difference between the series and shunt ohmmeters is by the scale of the meter. Figure 1-34 shows the scale of a series ohmmeter and the scale of a shunt ohmmeter.

Series and shunt ohmmeter scales - RF Cafe

Figure 1-34. - Series and shunt ohmmeter scales.

Figure 1-34(A) is the scale of a series ohmmeter. Notice "0" is on the right and "∞" is on the left. Figure 1-34(B) is the scale of a shunt ohmmeter. In the shunt ohmmeter "∞" is on the right and "0" is on the left. a shunt ohmmeter circuit is shown in figure 1-35.

In figure 1-35, R1  is a rheostat used to adjust the ∞ reading of the meter (full-scale deflection). R2, R3, and R4 are used to provide the R x 1, R x 10, and R x 100 ranges. Points a and B represent the meter leads. With no resistance connected between points a and B the meter has full-scale current and indicates

1-36

∞. If a resistance is connected between points a and B, it shunts some of the current from the meter movement and the meter movement reacts to this lower current. Since the scale of the meter is marked in ohms, the resistance of the shunting resistor (between points a and B) is indicated. Notice that the switch has an ofF position, as well as positions for R x 1, R x 10, and R x 100. This is provided to stop current flow and prevents the battery from being discharged while the meter is not being used.

A shunt ohmmeter with internal range resistors - RF Cafe

Figure 1-35. - a shunt ohmmeter with internal range resistors.

The shunt ohmmeter is connected to the circuit to be measured in the same way the series ohmmeter is connected. The only difference is that on the shunt ohmmeter the ∞ reading is adjusted, while on the series ohmmeter the 0 reading is adjusted. Shunt ohmmeters are not commonly used because they are limited generally to measuring resistances from 5 ohms to 400 ohms. If you use a shunt ohmmeter, be certain to switch it to the ofF position when you are finished using it.

Q46.    What are the two types of ohmmeters?

Q47.    What is the most obvious difference between the two types of ohmmeters?

Q48.    List the four safety precautions observed when using ohmmeters.

OHMMETER Safety PRECautionS

The following safety precautions and operating procedures for ohmmeters are the MINIMUM necessary to prevent injury and damage.

·   Be certain the circuit is deenergized and discharged before connecting an ohmmeter.

·   Do not apply power to a circuit while measuring resistance.

·   When you are finished using an ohmmeter, switch it to the ofF position if one is provided and remove the leads from the meter.

·   Always adjust the ohmmeter for 0 (or ∞ in shunt ohmmeter) after you change ranges before making the resistance measurement.

1-37

MEGOHMMETER

An ordinary ohmmeter cannot be used for measuring resistance of multimillions of ohms, such as in conductor insulation. To adequately test for insulation break down, it is necessary to use a much higher potential than is furnished by the battery of an ohmmeter. This potential is placed between the conductor and the outside surface of the insulation.

An instrument called a MEGOHMMETER (MEGGER) is used for these tests. The megger (fig. 1-36) is a portable instrument consisting of two primary elements: (1) a hand-driven dc generator, G, which supplies the high voltage for making the measurement, and (2) the instrument portion, which indicates the value of the resistance being measured. The instrument portion is of the opposed-coil type, as shown in figure 1-36(A). Coils a and b are mounted on the movable member c with a fixed relationship to each other, and are free to turn as a unit in a magnetic field. Coil b tends to move the pointer counterclockwise, and coil a tends to move the pointer clockwise.

A megger internal circuit - RF Cafe

Figure 1-36. - a megger internal circuit.

Coil a is connected in series with R3  and the unknown resistance, Rx, to be measured. The combination of coil, R3, and Rx  forms a direct series path between the positive (+) and negative (-) brushes of the dc generator. Coil b is connected in series with R2  and this combination is also connected across the generator. There are no restraining springs on the movable member of the instrument portion of the megger. Therefore, when the generator is not operated, the pointer floats freely and may come to rest at any position on the scale.

1-38

The guard ring intercepts leakage current. Any leakage currents intercepted are shunted to the negative side of the generator. They do not flow through coil a; therefore, they do not affect the meter reading.

If the test leads are open-circuited, no current flows in coil a. However, current flows internally through coil b, and deflects the pointer to infinity, which indicates a resistance too large to measure. When a resistance such as Rx  is connected between the test leads, current also flows in coil a, tending to move the pointer clockwise. At the same time, coil b still tends to move the pointer counterclockwise. Therefore, the moving element, composed of both coils and the pointer, comes to rest in a position at which the two forces are balanced. This position depends upon the value of the external resistance, which controls the relative amount of current in coil a. Because changes in voltage affect both coil a and coil b in the same proportion, the position of the moving system is independent of the voltage. If the test leads are short-circuited, the pointer rests at zero because the current in coil a is relatively large. The instrument is not damaged under these circumstances because the current is limited by R3.

The external view of one type of megger is shown in figure 1-36(B).

Navy meggers are usually rated at 500 volts. To avoid excessive test voltages, most meggers are equipped with friction clutches. When the generator is cranked faster than its rated speed, the clutch slips and the generator speed and output voltage are not permitted to exceed their rated values. When extremely high resistances-for example, 10,000 megohms or more-are to be measured, a high voltage is needed to cause sufficient current flow to actuate the meter movement. For extended ranges, a 1,000-volt generator is available.

When a megger is used, the generator voltage is present on the test leads. This voltage could be hazardous to you or to the equipment you are checking. Therefore, NEVER TOUCH The Test LEADS WHILE The MEGGER Is BEING useD and isolate the item you are checking from the equipment before using the megger.

Using the Megger

To use a megger to check wiring insulation, connect one test lead to the insulation and the other test lead to the conductor, after isolating the wiring from the equipment. Turn the hand crank until the slip clutch just begins to slip and note the meter reading. Normal insulations should read infinity. Any small resistance reading indicates the insulation is breaking down.

Megger Safety Precautions

When you use a megger, you could be injured or damage equipment you are working on if the following MINIMUM safety precautions are not observed.

·   use meggers on high-resistance measurements only (such as insulation measurements or to check two separate conductors on a cable).

·   Never touch the test leads while the handle is being cranked.

·   Deenergize and discharge the circuit completely before connecting a megger.

·   Disconnect the item being checked from other circuitry, if possible, before using a megger.

Q49.    What is the primary use of a megger?

1-39

Q50.    What is the procedure for using a megger to check the insulation of a conductor?

Q51.    What is a normal indication on a megger when checking insulation?

Q52.    List the four safety precautions observed when using a megger.

MULTIMETER

A MULTIMETER is the most common measuring device used in the Navy. The name multimeter comes from MULTIple METER, and that is exactly what a multimeter is. It is a dc ammeter, a dc voltmeter, an ac voltmeter, and an ohmmeter, all in one package. Figure 1-37 is a picture of a typical multimeter.

A typical multimeter - RF Cafe

Figure 1-37. - a typical multimeter.

1-30

NEETS Modules
- Matter, Energy, and Direct Current
- Alternating Current and Transformers
- Circuit Protection, Control, and Measurement
- Electrical Conductors, Wiring Techniques, and Schematic Reading
- Generators and Motors
- Electronic Emission, Tubes, and Power Supplies
- Solid-State Devices and Power Supplies
- Amplifiers
- Wave-Generation and Wave-Shaping Circuits
- Wave Propagation, Transmission Lines, and Antennas
- Microwave Principles
- Modulation Principles
- Introduction to Number Systems and Logic Circuits
- - Introduction to Microelectronics
- Principles of Synchros, Servos, and Gyros
- Introduction to Test Equipment
- Radio-Frequency Communications Principles
- Radar Principles
- The Technician's Handbook, Master Glossary
- Test Methods and Practices
- Introduction to Digital Computers
- Magnetic Recording
- Introduction to Fiber Optics
Note: Navy Electricity and Electronics Training Series (NEETS) content is U.S. Navy property in the public domain.